Endpoint:
particle size distribution (granulometry)
Type of information:
experimental study
Adequacy of study:
key study
Study period:
03 June - 8 June 2022
Reliability:
1 (reliable without restriction)
Rationale for reliability incl. deficiencies:
guideline study
Remarks:
Reliability 1
Qualifier:
according to guideline
Guideline:
other: CIPAC MT 187
Deviations:
no
Qualifier:
according to guideline
Guideline:
ISO 13320 (Particle size analysis - Laser diffraction methods)
Deviations:
no
GLP compliance:
yes (incl. QA statement)
Type of method:
Laser scattering/diffraction
Type of particle tested:
primary particle
Type of distribution:
volumetric distribution
Key result
Percentile:
other: D[4,3] – Mass Moment Mean – De Brouckere Mean Diameter
Mean:
224.44 µm
St. dev.:
1.3
Key result
Percentile:
D10
Mean:
34.006 µm
Remarks on result:
other: St. dev. not specified.
Key result
Percentile:
D50
Mean:
167.26
Remarks on result:
other: St. dev. not specified.
Key result
Percentile:
D90
Mean:
494.74
Remarks on result:
other: St. dev. not specified.

 

































































Particle sizeTest number and fraction size(µm)
12345Average
d10 [µm]34.6034.7734.3133.6232.7334.006
d50 [µm]168.4167.8167.6167.1165.4167.260
d90 [µm]487.5484.0490.2501.7510.3494.740
* d4,3 [µm]224.2222.6224.1225.2226.1224.440
Average d4,3, [µm]224.440
Standard deviation, [µm]1.313
Relative Standard Deviation, [%]0.585
Conclusions:
The distribution of particle size for the test item was determined to be: d10 = 34.006 µm, d50 = 167.260 µm, and d90 = 494.740 µm.
Executive summary:

The determination of particle size distribution for the test item was carried out in accordance with the method CIPAC MT 187 ( ISO 13320:2020) and following the Principles of GLP. The test was performed using Coulter LS 13 320 with ULM (Universal Liquid Model).A suspension of the test item of the adequate concentration was passed through a laser. The light scattered by the particles at various angles was measured by a multi-element detector and numerical values relating to the scattering pattern were then recorded for subsequent analysis. The percentiles for particle size distribution are: d10= 34.006 µm, d50= 167.260 µm, and d90= 494.740 µm. The mass moment mean (De Brouckere Mean Diameter) of the tested material is determined to be 224.440 µm.

Description of key information

Key study. Test method according to CIPAC MT 187 ( ISO 13320-1), GLP study. The percentiles for particle size distribution are: d10= 34.006 μm, d50= 167.260 μm, and d90= 494.740 μm. The mass moment mean (De Brouckere Mean Diameter) of the tested material is determined to be 224.440 μm.

Additional information