Registration Dossier

Data platform availability banner - registered substances factsheets

Please be aware that this old REACH registration data factsheet is no longer maintained; it remains frozen as of 19th May 2023.

The new ECHA CHEM database has been released by ECHA, and it now contains all REACH registration data. There are more details on the transition of ECHA's published data to ECHA CHEM here.

Diss Factsheets

Administrative data

Endpoint:
density, other
Type of information:
experimental study
Adequacy of study:
key study
Reliability:
1 (reliable without restriction)
Rationale for reliability incl. deficiencies:
guideline study

Data source

Reference
Reference Type:
study report
Title:
Unnamed
Report date:
2022

Materials and methods

Test guidelineopen allclose all
Qualifier:
according to guideline
Guideline:
OECD Guideline 109 (Density of Liquids and Solids)
Version / remarks:
02 October 2012
Deviations:
no
Qualifier:
according to guideline
Guideline:
EU Method A.3 (Relative Density)
Version / remarks:
No. 440/2008 of 30 May 2008
Deviations:
no
GLP compliance:
yes
Type of method:
air comparison pycnometer (for solids)

Test material

Constituent 1
Reference substance name:
Reaction mass of benzoyl chloride, toluoyl chloride and 2-methyl resorcinol
Molecular formula:
not applicable
IUPAC Name:
Reaction mass of benzoyl chloride, toluoyl chloride and 2-methyl resorcinol
Test material form:
solid: particulate/powder
Details on test material:
Appearance: beige powder

Results and discussion

Densityopen allclose all
Key result
Type:
density
Density:
1 250 kg/m³
Temp.:
20 °C
Type:
relative density
Density:
1.25
Temp.:
20 °C

Any other information on results incl. tables

Calibration
The pressure readings and the calculated volume for the reference and sample cells are shown in the following table:


Table 2










































 


Determination



 


P1


(psig)



 


P2


(psig)



 


P3


(psig)



 


P4


(psig)



Reference Cell Volume (cm3)



Sample Cell Volume (cm3)



1



17.145



6.486



17.204



8.520



90.620



148.923



2



17.172



6.502



16.986



8.415



90.859



149.102



3



17.083



6.471



17.194



8.519



90.986



149.212



Mean reference cell volume: 90.822 cm3


Mean sample cell volume: 149.079 cm3


As the reference cell volume and sample cell volume were within the defined tolerance (90.953 ± 1.0 cm3 and 149.225 ± 1.0 cm3 for the reference and sample cell volumes respectively), the gas comparison pycnometer was therefore considered acceptable for use.


Sample
The pressure readings, calculated volumes and density values obtained for the test item are shown in the following table:


Table 3





























































 


Determination



 


Replicate



P1(sample) (psig)



P2(sample) (psig)



Volume (cm3)



Density (kg/m3)



Mean Density (kg/m3)



 


1



A



17.090



7.160



23.121



1.25 x 103



 


1.25 x 103



B



17.160



7.191



23.171



1.25 x 103



C



17.193



7.203



23.116



1.25 x 103



 


2



A



17.069



7.111



21.895



1.28 x 103*



 


1.24 x 103



B



17.227



7.199



22.567



1.24 x 103



C



17.121



7.153



22.515



1.24 x 103



Mean density : 1.25 x 103 kg/m3
Relative density : 1.25
Temperature : 20.0 ± 0.5 °C

Applicant's summary and conclusion

Conclusions:
The density of the test item has been determined to be 1.25 x 10^3 kg/m3 at 20.0 ± 0.5 °C;
relative density: 1.25.